Quick Start-up and Shut-down Guide for Zeiss LSM 880 Confocal Microscope
INSTRUMENT STARTUP – Confocal Microscope
- Flip switches labeled 1, 2, and 3 in numerical order (1-2-3). Wait 6 seconds between each switch. Do not turn on or adjust any other switches.
- Start up computer and launch Zen software.
INSTRUMENT SHUTDOWN – Confocal Microscope (No User Scheduled After You)
Check the instrument calendar. If no session is scheduled within 1 hour of the end of your session, proceed with shutdown procedure.
- Clean the objective that you worked with
- Put the 10x objective in the working position, in the lowest z position
- Turn off Argon laser in software, and wait for fan to turn off before shutting down the instrument.
- Gather your data and shut down computer.
- Turn off instrument by flipping the switches used for start up in reverse numerical order (3-2-1). Do not turn off or adjust any other switches.
- Cover the microscope with the dust cover.
- Post a message on the slack channel to confirm that you have followed the protocol above before leaving the room.
- Special note to incubator users: please confirm that you have shut off the CO2. Let us know the CO2 level, and water level in the bottle.
INSTRUMENT STANDBY – Confocal Microscope (If User is Scheduled After You)
Check the instrument calendar. If a session is scheduled within 1 hour of the end of your session, proceed with the following procedure.
- Send a message in the Slack channel publically tagging the next user to see if they are still planning to come in at their scheduled time.
- Clean the objective that you worked with
- Put the 10x objective in the working position, in the lowest z position
- If the next user has confirmed that they will arrive within one hour, put the argon laser on standby and leave the system on for the next user.
- If you haven’t heard back from the next user, shut the system down following the INSTRUMENT SHUTDOWN protocol above.
- Post a message here on the slack channel to confirm that you have followed the protocol above before leaving the room.
- Special note to incubator users: please confirm that you have shut off the CO2. Let us know the CO2 level, and water level in the bottle.